Home / Latest recalls / 14V193

Wheeled Coach T3 Recall 14V193: Electrical System:Software; Equipment:Electrical

NHTSA campaign 14V193, reported April 16, 2014, 84 vehicles potentially affected.

Check whether this recall is open on your vehicle

The repair is normally free at an authorized dealer.

Official recall notice

Safety recall 14V193Wheeled Coach Industries, Inc.
Component
ELECTRICAL SYSTEM:SOFTWARE; EQUIPMENT:ELECTRICAL
Problem
Wheeled Coach Industries (Wheeled Coach) is recalling certain model year 2010 M2, 2013-2014 T1, T2, and T3, and 2014 Terrastar ambulances, equipped with ECX Timer part number EL00230 manufactured by ECX Electronics Corporation. The ECX timer may experience a momentary loss of power when starting the vehicle.
Risk
The momentary loss of power may cause the ECX timer to begin a five minute countdown, after which the timer would disconnect the power supplied to the electrical systems in the patient area of the vehicle, increasing the risk of injury to the occupants.
Remedy
Wheeled Coach will notify owners, and dealers will update the timer software, free of charge. Owners may contact Wheeled Coach at 1-800-628-8178.
Notes
Owners may also contact the National Highway Traffic Safety Administration Vehicle Safety Hotline at 1-888-327-4236 (TTY 1-800-424-9153), or go to www.safercar.gov.

Affected vehicles

Wheeled Coach AMBULANCE-M2 Chassis2010
Wheeled Coach T120132014
Wheeled Coach T220132014
Wheeled Coach T320132014
Wheeled Coach Terrastar2014